The cooling, low temperature, constant temperature and damp heat of the equipment provide the required cooling capacity. According to the different test states of the equipment, the refrigeration system will automatically start to provide cooling capacity for the corresponding test process, so as to meet the performance index of the equipment.
Certification :
Calibration Certificate(cost additional)Price :
NegotiableDelivery Time :
30 daysPayment Terms :
L/C, T/TBrand Name :
HCTEModel Number :
HCTE0076High and Low Temperature Alternating Damp Heat Test Equipment
Products Introduction:
The cooling, low temperature, constant temperature and damp heat of the equipment provide the required cooling capacity. According to the different test states of the equipment, the refrigeration system will automatically start to provide cooling capacity for the corresponding test process, so as to meet the performance index of the equipment.
Standard:
IEC60068
Technical Parameter:
Model |
HCTE0076 |
Volume and size |
150L(can place 3 products) |
Chamber size(D*W*H mm) |
500×500×600 |
External dimensions(D*W*H mm) |
About 1000×1100×1600 |
Ambient temperature |
5~35℃ |
relative humidity |
≤85%RH |
pressure |
Air quality |
Specimen Limit
|
This test equipment prohibits: Testing and storage of flammable, explosive and volatile material samples Testing and storage of corrosive substance specimens Testing or storage of biological samples Test and storage of samples of strong electromagnetic emission sources |
Performance indicators Test environmental conditions |
: no high-concentration dust and corrosive gases, and cannot be tested and stored in flammable and explosive gases |
Tap water |
Tap water requirements: Water pipe: 1/2″ Water pressure: 0.1~0.3 Mpa |
Temperature range |
-70℃ ~+200℃ |
Humidity range |
30%~95%RH |
Temperature fluctuation |
±0.5℃ |
Humidity fluctuation |
±2.5%RH |
Temperature deviation |
≤±2.0℃ |
Humidity deviation |
≤± 2%R.h(>75%), ≤± 2.5%R.h.(<75%) |
Heating time |
-70℃~ 200℃≤43min |
Cooling time |
+200 ℃~ -70℃≤90min |
Noise |
60dB |
Test function |
Constant temperature fixed value, constant damp heat, heating, cooling, high and low temperature cycle test |